6. Lubrication
CAUTION
Supply PANA SPRAY Plus or/and Care3 Plus after each use and/or before autoclaving.
PANA SPRAY Plus
1. Push E-type spray nozzle attachment over the PANA SPRAY Plus
nozzle until it firmly seats.
2. Remove the file from handpiece.
3. Shake the can 3-4 times to well mix lubricant and propellant.
4. Insert the E-type spray nozzle in the rear of the handpiece and spray
for approximately 2-3 seconds until the oil comes out of the
handpiece head.
8
E-Type Spray Nozzle
PANA SPRAY Plus
Handpiece
PANA SPRAY Plus
Fig. 8