1.8. TROUBLE SHOOTING

Problem
The power cord is connected to the
mains, the electrical switch S is ON
and the green light does not turn on.
Water
supply
distillation does not happen
In the event of overheating that
activates the safety mechanism.
Note:
It will not be allowed to repair any water still that is not properly cleaned.
Measure and check that the voltage in the network is 230 V.
(110 V for 10DESA1041 and 380 v for 10DESA0120)
Check the status of the switch.
is
adequate
and
Following this order:
Check level switch N. The circuit closes upwards and activates
contactor C1. Replace if necessary.
There is current supply to the input of contactor C1 and it does
not work: replace it.
There is current supply to the resistance: R, R1, R2, R3 and it
does not heat. Replace the resistance.
Press the button RESET.
Action
12