SETTING THE COMPRESSOR DELAY

The delay time before the compressor starts can be changed to reduce the load on the power line and to
facilitate the start-up (reset) of the freezer during recovery after power failure.
 Setting range of delay time: between 3 and 15 minutes, in 1-minute intervals
 Initial setting (factory setting): 3 minutes
Note:
There is no need to change the delay time when the capacity of the power source is adequate.
For example, to change the setting of the compressor delay from 3 minutes (initial setting) to 4 minutes
(The following is only an example. Choose the setting according to the desired start delay.)
Operation
1
Press the up arrow key for 5
2
seconds.
3 Press the up arrow key 5 times.
4 Press the "Set" key once.
Press the up arrow key (as many
5
times as needed to reach the
desired figure).
6 Press the "Set" key once.
The unit will return from setting mode to temperature display mode automatically after 90 seconds if no
key is operated (auto-return function). In this case, the setting is not changed..
26
Key operated
The current chamber temperature is
----
displayed.
The current chamber temperature
changes to function code F00 and
the right-hand figure (0) blinks.
The display changes from F00 to F05.
The current setting (003) is displayed
SET
and the right-hand figure (3) blinks.
The display changes from 003 to 004.
The new setting is memorized and
SET
the current chamber temperature is
displayed.
Display after operation