Gima 35132 Manual De Uso página 146

Ocultar thumbs Ver también para 35132:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Recommended separation distances between portable and mobile RF
communications equipment and The equipment or system-
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Vital Signs Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the equipment or system can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the equipment or system as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where
p is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2:These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
User Manual for Vital Signs Monitor
Table 4
Separation distance according to frequency of transmitter
m
150kHz to 80MHz
0.12
0.38
1.2
3.8
12
80MHz to 800MHz
0.12
0.38
1.2
3.8
12
55
80MHz to 2,5GHz
0.23
0.73
2.3
7.3
23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pc-900plus snet

Tabla de contenido