5

Adjusting the settings

To check and, if necessary, change the instrument
settings:
1.
Tap the title bar to open the menu (see
figure 2).
If the display is locked, tap the
i
symbol.
Pressing the green Start key will
close the menu at any time and
display the basic view.
Use the scroll bar on the far right of
the display to show further menu
options.
Measuring
conditions
Traversing length
Number N of
sampling lengths
Pre-travel/post-travel
Ls filter
Mahr GmbH, MarSurf PS 10
Tap
Lt = 1.5 mm
Fixed traversing length with half pre-travel/post-travel
Lt = 4.8 mm
(without pre-travel/post-travel: 1.25 mm, 4.0 mm,
Lt = 15.0 mm
12.5 mm).
Lt = N x Lc
Traversing length = number n of sampling lengths x
length of cutoff.
Traversing length freely defined (e.g. Lt = 16 mm for
Lt
mm
MOTIF).
Automatic
The traversing length is set automatically during the
measurement depending on the magnitude of the
value for RSm or Rz (as per DIN EN ISO 4288 and
ASME B46. 1 ).
One-off display in title bar "Lt = Automatic".
N 5
Number N of sampling lengths (as per ISO, JIS,
ANSI/ASME).
Entry of N = 1 to 16 (only for "Lt = N x Lc").
N
Half
Traversing length includes a pre-travel and a post-
travel measuring half the length of the cutoff.
None
Traversing length without pre-travel/post-travel.
Automatic
Profile is automatically filtered after the measurement
with a cutoff wavelength of 2.5 µm or 8.0 µm de-
pending on the traversing length.
2.5 µm
Ls filter with a fixed value.
8.0 µm
Off
Ls filter is disabled.
2.
Tap the relevant menu item.
3.
Adjust the required settings:
Option is not selected.
Option is selected.
Function is not enabled.
Function is enabled.
Input box.
4.
Tap OK.
to change.