3. Align the black mark on the translucent tubing with the center
pinch plunger of the Rinse Valve.
4. Place the translucent tubing within the Pinch Valve slot and pull
straight down until the translucent tubing stretches enough to slip
completely down into the Rinse Valve slot.
5. Being careful not to twist the Rinse Tubing, guide tubing into
a "U" shape and place the plastic connectors that connect the
translucent tubing to the gray/blue tubing into the recesses above
the Rinse Valve.
6. Connect the Saline Bag Spike (left side of installed Rinse Tube
Set) to a bag of sterile saline and place the bag on the upper
shelf of the vacuum system.
7. Install the Vacuum Canister Tube Set per the Vacuum Tube Set
instructions.
8. When ready, the Biopsy Device Tube Set can be connected to
the Rinse Tube Set (right side of installed Rinse Tube Set) and
passed through the opening at the top of the VS 3000 door when
closed.
9. At the conclusion of the procedure, remove the Rinse Tube Set
and dispose of properly.
Rinse
Tubing
Set
Rinse
Pinch
Valve