PARTS REPLACEMENT
Remove the nut from pendant cord,
and install the nut on the new pendant
cable. Make sure the ribbed end of
the nut is toward the connector end of
the pendant cable.
Disconnect the pendant cable from
the Allen Advance Table connector.
Replacement
Make sure the O-ring is installed in the Allen Advance Table
connector.
Align the tab on the connector end of the new pendant cable with the
tab on the inside of the Allen Advance Table connector.
Connect the pendant cable to the Allen Advance Table connector.
Screw the nut on to the Allen Advance Table connector.
Plug the Allen Advance Table in, and turn it on.
Make sure all pendant functions operate correctly.
FUSES
Removal
Make sure the brakes are locked (the locked indicator on
the Allen Advance Table display is green).
WARNING:
Failure to remove power from the Allen
Advance Table could cause patient fall, injury, or equipment
damage.