Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The A-71100-XX/ A-71101-XX are intended for use in the electromagnetic environment
specified below. The customer or the user of the A-71100-XX/A‑ 71101-XX should make
sure it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration—Electromagnetic Immunity
The A-71100-XX/ A-71101-XX are intended for use in the electromagnetic environment
specified below. The customer or the user of the A-71100-XX/A‑ 71101-XX should make
sure it is used in such an environment.
Immunity Test
Static Electricity
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Allen® Advance Table User Manual (D-720761 Ver D)
Compliance
Electromagnetic Environment—Guidance
Group 1
The A-71100-XX/A-71101-XX use RF energy
only for its internal functions. Therefore, its RF
emissions are low and are not likely to cause any
interference in nearby electronic equipment.
Class A
The A-71100-XX/A-71101-SCH are suitable for
use in all establishments, other than domestic,
and those directly connected to the public low-
voltage power supply network that supplies
Class A
buildings used for domestic purposes.
Complies
IEC 60601
Test Level
± 6 kV Contact
± 8 kV Air
± 2 kV (power line)
± 1 kV
(Input/Output line)
± 1 kV (line – line)
± 2 kV (line – earth)
Electromagnetic
Compliance
Environment—
Level
Guidance
± 6 kV Contact
Floors should be wood,
± 8 kV Air
concrete, or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
± 2 kV (power line)
Mains power quality
± 1 kV
should be that of a
(Input/Output line)
typical commercial or
hospital environment.
± 1 kV (line – line)
Mains power quality
± 2 kV (line –
should be that of a
earth)
typical commercial or
hospital environment.
Specifications
57