Cleaning: Automated
1. Brush
• Brush both ends of any lumens a minimum of five times, using an
appropriate bottle brush.
2. Rinse
• Rinse the device with treated water at ambient temperature until
there is no visible detergent residue. Continue to rinse for a
minimum of 30 seconds after all detergent residue has been
removed.
• Place the device in the washer on an incline to facilitate drainage.
3. Automated wash
• Program the washer using the following parameters:
Phase
Pre Wash
Enzyme
Wash
Wash 1
Rinse 1
Dry Phase
* If necessary, use pressurized air to aid in drying. Visually inspect each
device for cleanliness.
Disinfection (optional)
1.
Disinfect the device by submerging it in a disinfection agent. Follow
the disinfection agent manufacturer's recommended concentrations,
temperatures, and exposure times.
2.
Thoroughly rinse and flush all parts and lumens with running,
demineralized water to remove the disinfectant.
3.
Dry all parts with a lint-free towel immediately after rinsing.
EN-32
Recirculation
Time
2 minutes
2 minutes
2 minutes
2 minutes
7 minutes
Water
Detergent Type
Temperature
and Concentration
(If applicable)
Cold
N/A
Hot
Enzymatic
Detergent
Set Point
Regular Detergent
(66° C)
Hot
N/A
115° C
N/A