Assembly
WARNING!
Danger of cutting when using the hose cutter.
`
`
Keep hands away from the cutting area.
1. Screw the sample box  onto the instrument
finger-tight.
2. Guide the suction hose  through the open clam-
ping screw  about 1 cm into the sample box.
3. Screw on the clamping screw finger-tight and fix the
suction hose in place.
4. Place the free end of the suction hose in the liquid
you are sampling. Ensure that the suction hose is
not kinked.
5. Place the hose weight  on the suction hose
depending on the location where it will be used and
tighten the union nut. The hose weight improves the
stability of the suction hose.
NOTE!
Possible loss of the hose weight during use.
`
`
Screw on the hose weight tightly with the cap
nut.
`
`
Verify that the hose weight is attached correctly
before use.
Troubleshooting
Problem
Pump is not generating a
negative pressure
Accessories
Type
Sample box (PFA)
Sample box (PFA)
Suction hose (PTFE, on roll)
PTFE MiniSampler adapter
More accessories can be found at www.buerkle.de
To eliminate the error
Repeat assembly steps 1 to 4.
Seal or valve faulty (contact dealer or manufacturer).
Safe sampling
WARNING!
Bodily injury due to escaping media when over-
filling the sample box. Liquids can be aspirated
into the piston and can escape at the back of the
instrument.
`
`
Do not overfill the sample box.
`
`
Wear the appropriate protective clothing.
1. Activate the handle  and pump liquid into the
sample box.
2. Screw off the sample box and seal it.
WARNING!
Bodily injury due to escaping media when screwing
off the sample box and pulling the suction hose
out of the instrument.
`
`
Dismount the instrument carefully.
`
`
Wear the appropriate protective clothing.
After taking the sample
1. Open the clamping screw and remove the suction
hose.
2. Rinse the suction hose (together with the hose
weight), sample box and instrument with a pH-neu-
tral liquid.
Specification
90 ml
180 ml
Ø 6x8 mm, length 5 m
for glass thread GL45
English
Number
Item No.
1
6205-0090
1
6205-0180
1
8804-0608
1
5305-0209

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

5305-0110

Tabla de contenido