pico BLED
1 ) Remove the handpiece from the coupling and loosen the Taper
Ring to remove. ( Fig. 7 )
2 ) Pull out the joint then remove the Non-Retraction Valve Pull out
the old Non-Retraction Valve with thin stick. ( Fig. 10 )
3 ) Insert New Non-Retraction Valve securely.
4 ) Mount the Joint into the handpiece. Make certain the pipe and
joint together. ( Fig. 11 )
5 ) Tighten the Taper Ring tight.
CAUTION
Make certain that the Taper Ring is firmly tightened. If the Taper Ring is loose water and air leakage could occur.
Joint
Non-Retraction Valve
Fig. 10
Handpiece
Joint
Align
Fig. 11
11