PERFORMANCE SPECIFICATIONS
Table information provides A
clinical study.
SaO
versus error (SpO
2
2
values measured using the YI sensors with Masimo SET Oximetry Technology in a
rms
RANGE
90–100%
80–90%
70–80%
OVERALL CLAIMED ACCURACY VALUE
70–100%
- SaO
) with linear regression fit and upper 95% and lower 95% limits of agreement.
2
Reusable Sensor
MEASURED ARMS VALUES
A
1.45%
1.22%
1.41%
1.36%
140
rms
9024C-eIFU-0518