Hach NA5600 sc Instrucciones Para El Usuario página 18

Ocultar thumbs Ver también para NA5600 sc:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 73
4. Select an option.
Option
SELECT SOURCE
SET LOW VALUE
SET HIGH VALUE
SET TRANSFER
SET FILTER
SET HART
ADDRESS
5. Push back.
6. Select DIAG/TEST > ERROR HOLD MODE.
7. Select an option.
Option
Description
HOLD(default) Holds the outputs at the last known value when an error occurs or
measurements are stopped (e.g., calibration, rinse, reactivation or grab sample
measurement).
TRANSFER
Sets the outputs to the transfer value when an error occurs or measurements
are stopped (e.g., calibration, rinse, reactivation or grab sample
measurement).
Section 6 Diagnostics
Do tests on the module and see module information as follows:
1. Push menu, then select SETUP SYSTEM > SETUP NETWORK > DIAG/TEST.
2. Select an option.
Option
OUTPUT CAL
HOLD OUTPUTS
18 English
Description
NONE (default)—Sets the output to disabled. [analyzer name]—Sets the
output to enabled. Selects the channel represented at the output.
Note: The channels that are followed by an asterisk (*) are disabled.
Sets the low measurement value that is represented as 4 mA at the
analog output (default: 0.000).
Sets the high measurement value that is represented as 20 mA at the
analog output (default: 1.000).
Sets the value at the selected output when an error occurs if the ERROR
HOLD MODE setting is set to TRANSFER OUTPUTS (default: 4 mA).
Sets the amount of time for analog output averaging (default: 0 seconds).
For example, if the value is set to 30 seconds, the value of the analog
outputs is updated every 30 seconds and the value is the average of the
analog output values during the previous 30 seconds.
Sets the HART address for the selected output channel. Range 0 (default)
to 63.
Description
Calibrates the low (4 mA) and high (20 mA) values for each of the three
4-20 mA outputs (A, B, C).
Adjust the DAC counts until a connected multimeter reads the
applicable output value. The default DAC counts are:
• Output A—2807 for 4 mA, 14046 for 20 mA
• Output B—2797 for 4 mA, 14109 for 20 mA
• Output C—2641 for 4 mA, 13294 for 20 mA
HOLD—Holds the selected output at the last known value. TRANSFER
—Sets the selected output to the transfer value. RELEASE (default)—
Lets the selected output continue to represent the measured value.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Na9600 sc

Tabla de contenido