6.3. Nasal Cannula

Only use a nasal cannula with the following specifications:
• 7ft (2.1 m) or 25ft (7.6 m) long
• High flow
• Crush resistant
• Large internal diameter bore
• Straight non-tapered tips
• Suitable for up to 15 litres per minute (lpm) at a max. pressure of 3.6 psi
• Meets substance compatibility of IEC/EN 60601-1
EN
CAUTION: Only use approved accessories with this device. Refer to the approved accessories
guide for a complete list of accessories and cannula approved for use with this device. Using
unapproved accessories or cannula may impair the performance of this device, including flow
rate or oxygen purity.
Contact your distributor for updated information and accessories or if additional, optional, or
replacement accessories are needed.

6.4. Pull Cart

When using the device with a pull cart, attach and
secure the concentrator with the straps as shown
in Fig. 4. The handle can be pulled out and adjusted
for comfort.
IMPORTANT: It is recommended that patients use the
pull cart to transport the device whenever possible.

Operating Zen-O™

7 .
IMPORTANT: Read Chapter 3. Safety Instructions be-
fore using this device.
Zen-O™ Portable Oxygen Concentrator is designed
for ease of use, with all functions accessed through
just a few keys on the control panel.
The device should be carried in its carry bag, placed
on a cart and used when positioned upright on a
table or on the floor while in the carry bag. The pa-
tient should be within the recommended cannula
length during use.
IMPORTANT: Except during startup and shutdown
sequences, the backlight on the display screen will re-
Fig. 4
main off. Pressing any key will turn the backlight on
briefly. The backlight will also remain activated during an un-muted alarm condition.