OPERATOR'S MANUAL
TP702 T/PUMP

EMC INFORMATION

The TP702 is suitable for use in the electromagnetic environment specified below.
The customer or the user of the TP702 should assure that it is used in such an environment.
Immunity test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrical fast Transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
The TP702 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
the TP702 can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the TP702 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
www.gaymar.com
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5%Ut(95% dipUt) for
0,5 cycle
40%Ut (60% dop in Ut) for 5
cycles
70% Ut (30% dip in Ut) for 25
cycles
<5% Ut (>95% dip in Ut) for
5 sec.
3 A/m
Recommended separation distances between
portable and mobile RF communications equipment and the TP702.
150 kHz to 80 MHz
P
d = 1,2
0,12
0,38
1,2
3,8
12
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5%Ut(95% dipUt) for 0,5
cycle
40%Ut (60% dop in Ut)for 5
cycles
70% Ut (30% dip in Ut) for 25
cycles
<5% Ut (>95% dip in Ut) for
5 sec.
3 A/m
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
P
d = 1,2
0,12
0,38
1,2
3,8
12
EMC INFORMATION
Electromagnetic environment guidance
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital environment
Mains power quality is that of a typical
commercial and/or hospital environment
Mains power quality should be that
of a typical commercial or hospital
TP702
environment. If the user of the
requires continued operation during power
mains interruptions, it is recommended
that the device be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial and/or hospital
environment.
800 MHz to 2,5 GHz
P
d = 2,3
0,23
0,73
2,3
7,3
23
19