Table 3: Guidance and manufacturer's declaration – electromagnetic immunit
The bosolux auris is intended for use in the electromagnetic environment specifi ed below. The
customer or the user of the bosolux auris should assure that it is used in such an environment.
Immunity
IEC 60601
test
test level
Conducted
3 Vrms
RF IEC
150 kHz to
61000-4-6
80MHz
Radiated
3 Vrms
RF
80MHz to
IEC
2.5GHz
61000-4-3
Com-
Electromagnetic environment-guidance
pliance
level
3V
Portable and mobile RF communications equipment
should be used no closer to any part of the SYSTEM,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
3V/m
Recommended separation distance
Where P is the maximum output power rating of the
transmitter in watts(W) according to the transmitter manu-
facturer and d is the recommended separation distance
in metres (m).
Field strengths from fi xed RF transmitters, as determined
by an electromagnetic site survey, should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment
marked wigh the following symbol:
- 41 -
80MHz to 800MHz
800MHz to 2.5GHz