COVER REPLACEMENT

Tools Required:
None
PROCEDURE
1.
Raise the bed height to the full up position.
2.
Lower the fowler and gatch sections to the full down positions.
3.
Unzip the cover. Start at the foot end patient right corner of the support surface and stop at the head end patient right
corner.
4.
Fold the top of the cover to the patient's right side and then remove the foam crib assembly from the bed and set
aside.
5.
Discard the old cover.
6.
Place the new cover, unzipped and open, on the bed with the black bottom cover on the litter and the top cover folded
over the patient's right side of the bed.
7.
Carefully place the foam assembly on top of the bottom part of the cover so it aligns with the cover.
NOTE! Maintain the correct "Head-Foot" orientation between the new cover and the foam assembly.
8.
Fold the top cover over the top of the foam crib assembly so it aligns with the foam crib assembly.
9.
Zip the cover to close. Start at the head end patient right corner and stop at the foot end patient right corner.
10.
Verify proper operation of the unit before returning it to service.

FIRE BARRIER REPLACEMENT

Tools Required:
None
PROCEDURE
1.
Raise the stretcher height to the full up position.
2.
Lower the fowler and gatch sections to the full down positions.
3.
Unzip the cover. Start at the foot end patient right corner of the support surface and stop at the head end patient
right corner.
4.
Fold the top of the cover to the patient's right side.
5.
Starting at the foot end, roll up the fire barrier on the foam crib assembly
NOTE! Work from side to side, a little at a time, to move the fire barrier to the top of the support surface.
6.
Discard the old fire barrier.
7.
Starting at the head end, roll the new fire barrier down and slide over the foam crib assembly.
NOTE! Align the fire barrier on the foam crib before sliding over the foam crib assembly.
8.
Carefully slide the fire barrier down the foam crib assembly, working from side to side, to make sure that the fire
barrier is tight on the foam crib assembly.
9.
Align the foam crib assembly on top of the bottom part of the cover.
NOTE! Spread the excess fire barrier material equally below the foam crib assembly at the foot end.
10. Fold and align the top cover over the top of the foam crib assembly.
11. Zip the cover to close. Start at the head end patient right corner and stop at the foot end patient right corner.
12. Verify proper operation before returning the product to service.