CAUTION for Bur Usage
Heat Generation Problem
• Bur position is improper or it is not mounted.
--> Mount the bur correctly.
• During operation, Bur Lock Ring had been turned.
--> Do not turn the Bur Lock Ring while the handpiece is
rotating.
Bur Jump Out Problem
• Incorrect mount of the bur/bur stopper.
--> Mount the bur/bur stopper firmly.
• During operation, Bur Lock Ring had been turned.
--> Do not turn the Bur Lock Ring while the handpiece is
rotating.
4
• Mount short shank contra-angle bur.
--> Do not mount this bur.
Bur Breakage Problem
• Mount damage or sub-standard burs.
--> Do not mount those burs.
• Exceed drive motor speed or manufacture's recommended bur
speed.
--> Do not exceed drive motor speed or manufacture's
recommended bur speed.
3. External Irrigation Nozzle
Connect the irrigation tube to the external irrigation nozzle firmly.
4. Maintenance
CAUTION
After each patient, be sure to keep the handpiece after cleaning, lubrication, and sterilization. If the blood is left on the handpiece, it can be
clotted causing rusty.
OM-Z0135E.
Refer to the Maintenance Guide
This handpieces can be autoclavable up to Max.135°C.
This handpieces can be cleaned and disinfected with a Thermo-Disinfector.
Irrigation Tube
External Irrigation Nozzle
Fig. 5
*Cleaning
Clean the external irrigation nozzle inside with clean water.
5