STEP 3:
PLACE ARM IN DEVICE & ADJUST TO START
POSITION FOR JAS SESSION
• Attach clip-on stand to base. Rest stand on knee or table
for stability while fitting device.
• Place arm in device; therapy knob should be on the inner
side.
• Rotate therapy knob to adjust device angle until it matches
your comfortable end range of motion.
• Align tip of elbow over fulcrum marker (see arrow).
NOTE: Forearm position (palm up or down) is optional.
Choose position of comfort.
STEP 4:
SECURE UPPER ARM CUFF STRAPS
• Feed straps through D – rings on upper arm cuff and secure
so that cuff fits comfortably snug.
STEP 5:
ADJUST POSITION OF FOREARM CUFF & SECURE STRAPS
• Loosen knob under forearm cuff, and slide cuff forward or
back to comfortable position. Tighten knob.
• Secure velcro straps on forearm cuff comfortably snug.
• Remove clip-on stand.
• Begin treatment by rotating therapy knob as outlined on
.
back page
IMPORTANT FOR FLEXION USE
When working in flexion beyond 95°, bunching in upper
cuff or colliding of both cuffs may occur, preventing further
flexion. To remedy:
a: If using for Flexion only: Trim off front portion of upper cuff.
b: If using for both Flexion & Extension: Loosen upper arm
straps & move cuff out of way during flexion sessions.