Equipment Setup And Use; Prior To Use - Stryker F-SFFFBER Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 620
INSTRUCTIONS FOR USE

3. Equipment Setup and Use:

Prior to use:

a. Inspect the product looking for any visible damage or sharp edges that could
be caused by a drop or impact during storage.
b. Make sure product has been properly cleaned and disinfected and wiped dry
prior to each use.
c. Position and orient table in the room. Spine
System will extend approximately 5' (1525
mm) from end of OR table.
d. Remove head section if attaching to back
section.
OR
Remove lower leg section if attaching to
seat section.
e. Place floor lock pads under table, then
engage floor locks. Verify floor locks are
engaged by tugging on floor lock pads.
Ensure OR table is rated for patient weight at connection
(If this information is not available it should be tested prior
to loading a patient).
Document Number: 80028282
Version: B
Page 15
5'(1525 mm)
Issue Date: 15 APR 2020
Ref Blank Template: 80025118 Ver. E

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

80028282

Tabla de contenido