3. Equipment Setup and Use:
Prior to use:
a. Inspect the product looking for any visible damage or sharp edges that could
be caused by a drop or impact during storage.
b. Make sure product has been properly cleaned and disinfected and wiped dry
prior to each use.
c. Position and orient table in the room. Spine
System will extend approximately 5' (1525
mm) from end of OR table.
d. Remove head section if attaching to back
section.
OR
Remove lower leg section if attaching to
seat section.
e. Place floor lock pads under table, then
engage floor locks. Verify floor locks are
engaged by tugging on floor lock pads.
Ensure OR table is rated for patient weight at connection
(If this information is not available it should be tested prior
to loading a patient).