Do not realise measurements if the battery compartment is
open.
Do not commence measurements if the test leads are con-
nected to the meter.
In order to realise a measurement of current, it is necessary to
realise the following actions:
Set the Function switch to the 1000A or 660A range. If the ap-
prox. range of the measurement is not known, select the high-
est range then move to the lower ranges if necessary,
Press the DC ZERO button to zero the meter display,
Press the trigger to open jaw. Fully enclose only one wire. For
optimum results, center the wire in the jaw,
Read the result of the measurement on the display.
During measurements of the current make sure the clamp is
properly placed. Otherwise the results of the measurements
will not be exact. The most exact result we will get if the wire
is placed in the middle of clamp.
WARNING:
WARNING:
Attention: