Other technical data
a) Measurement category ...........III 600V according to EN 61010-1
b) Clamp jaw opening......................34mm (1.34") approximately
c) Internal diameters of clamp...................36x52mm (1,42x2,05")
d) Display..........................................6600 counts backlit LCD
e) Continuity check..............threshold 40; Test current < 0.5mA
f)
Diode test.................................test current of 0.3mA typical;
g) Low Battery indication................................'BAT' is displayed
h) Over-range indication..................................'OL' is displayed
i)
Measurement rate...................2 readings per second, nominal
j)
Inrush.........................integration time 100ms, sampling 10ms
k) Temperature sensor..............................type K thermocouple
l)
Input Impedance.................................10M (VDC and VAC)
m) AC bandwidth............................50 to 400Hz (AAC and VAC)
n) Operating temperature..................5
o) Storage temperature...................-20
p) Operating humidity......................max 80% up to 31
q) Storage humidity......................................................<80%
r)
Operating altitude...........................2000m (7000ft) maximum
s) Battery........................................................one 9V battery
t)
Auto power OFF......................after approximately 25 minutes
u) Dimensions........................229 x 80 x 49mm (9.0 x 3.1 x 2.0")
v) Weight.....................................................................303g
For indoor use and in accordance with the requirements for
double insulation to IEC1010-1 (1995): EN61010-1 (1995)
Overvoltage Category III 600V, Pollution Degree 2.
42
open circuit voltage < 3VDC typical
decreasing linearly to 50% at 40
Attention:
CMP-1006 – USER MANUAL
o
o
C to 40
C (41
o
o
C to 60
C (-4
o
o
F to 104
F)
o
o
F to 140
F)
o
o
C F (87
)
o
o
C (104
F)