Limit 94
OPERATION MANUAL
English
Overview
LIMIT 94 professional non-contact infrared thermometer can determine surface temperature through
measuring infrared energy radiated by the target surface. Compact and lightweight design easy to carry.
Warning:
To avoid eyes injury, do not look directly in to the laser beam.
CAUTION
!
LASER
2
Features
• Dual laser beams
• User selectable ºC or ºF
• Compact and ergonomic design
• Distance to spot ratio 12:1
• Automatic Data Hold
• Inverted display
• MAX/MIN function
Specifications Limit 94
IR measurement range:
IR accuracy:
D:S ratio:
Emissivity:
Response time:
Spectral response:
Temperature resolution:
IR repeatability:
Laser beam qty:
Laser type:
Laser wavelength:
Laser power:
Battery type:
Battery life:
Product size:
Weight:
Operating temperature:
Operating humidity:
Operating altitude:
Storage altitude:
Protection level:
Positions
1. Measurement trigger
2. Laser On / Off key
3. °C / °F and buzzer on / off button
4. Max / Min key
5. LCD display
6. Battery cover
-50°C ~ 380ºC (-58°F ~ 716°F)
-50°C ~ 20°C (-58°F ~ 68°F): ±2.5°C (4.5°F)
20°C ~ 380°C (68°F ~716°F): ±1.0% ±1.0°C (1.8°F)
12:1
0.95
150 ms
8 um ~ 14 um
0.1 °C (0.1°F)
-50°C ~ 20°C (-58°F ~ 68°F): ±1.3°C (2.3°F)
20°C ~ 380°C (68°F ~ 716°F): ±0.5% ±0.5°C
2
CLASS II
630 nm ~ 670 nm
<1 mW
9 V (1604A, 6LR61)
≤ 10 h
131 mm × 96 mm × 35 mm
130 g
0°C ~ 50°C (32°F ~ 122°F)
< 90% RH (non-condensing)
2000 m
12000 m
IP4x
5
4
1
English
3
2
1
GB
6