• Place the RV 10.8001 condenser seal (U) in the condenser
receptacle and fit the glassware to the device according to the
assembly instructions. (Fig. 16, 17 and 18)
U
Mount the condenser
• Place the cap nut into the condenser, and then put the spring
ring to the flange of the condenser (V).
• Place the condenser on the rotary drive (W) and tighten the cap
nut by hand. (Fig.17)
V
W
• Fix the receiving flask and hose connectors as shown. See also
operating manual of glassware. (Fig.18)
Fig.16
Mounting the washer
Note: Pay attention to the
correct position of the washer.
Removing the condenser
• Use the ring spanner provided to loosen union nuts that are
tightly fitted.
• Loosen the union nut by turning anticlockwise.
Description of special condensers
(see products/accessories at www.ika.com)
• RV 10.3 Vertical-intensive condenser with manifold
Fig.17
This vertical-intensive condenser features a double jacket design
for particularly efficient condensation.
Also available with coating (RV 10.30)
• RV 10.4 Dry ice condenser
Dry ice condenser for distilling low-boiling solvents. Cooling
by dry ice, no cooling water required. Maximum condensation
thanks to low temperatures.
Also available with coating (RV 10.40)
Fig.18
Drainage
washer
Fig.19a
Fig.19b