Relative Humidity Item
Current/
110V ± 10V
Voltage Rating
power outlets
220V ± 20V
power outlets
Weight with Cable Cover
(approximate)
Unit Dimensions (W x H x D)
VESA Mounting Interface
Dimensions
Accessory Mounting Interface
Dimensions

Classification and Approvals

Class 1 Equipment
Medical equipment with respect to electric shock, fire, and mechanical hazards only in
accordance with ANSI/AAMI ES60601-1 and CAN/CSA C22.2 No. 60601.1.
IP23: Protection against access to hazardous parts from fingers or similar objects,
protection from ingress of spraying water (less than 60 ° from vertical)
Continuous Operation

Compliance

FCC Regulations: FCC Part 15 Class B
FCC Identifier: QVXAMM320ES
Canadian Notice: CAN ICES-3 (B)/NMB-3(B)
CE medical device class I
Note: Please contact your local Stryker Endoscopy sales representative for information on
changes and new products.
Notice to the user and/or patient: Any serious incident that has occurred in relation to
the device should be reported to the manufacturer and the competent authority of the
Member State in which the user and/or patient is established.
Description
Select a power supply cord that is UL Listed and C.S.A
Certified, type SJT or SVT, 3 – conductor, 18 AWG,
terminated in a molded on hospital grade plug cap
rated 110 V ± 10 V, 15 A, with a minimum length of six
feet.
Select a power supply cord that is internationally
harmonized and marked "<HAR>", 3 – conductor,
0.75 mm^2 minimum wire, rated 220 V ± 20 V, 10 A
with a PVC insulated jacket. The cord must have a
molded on plug cap rated 220 V ± 20 V, 10 A. The cord
and plug cap must be suitable for medical use.
23.1 lb (10.5 kg)
756.7 x 453 x 77.2 mm
29.7 x 17.8 x 3 in
VESA 100 x 100 mm
VESA 200 x 100 mm
75 x 75 mm (x 2)
EN-23

Publicidad