I I n n s s t t a a l l l l i i n n g g t t h h e e h h e e a a d d e e x x t t e e n n s s i i o o n n w w i i t t h h p p i i l l l l o o w w
You can install the head extension assembly on the backrest to provide head end support.
To attach the pillow to the head extension, place the support into the flap on the bottom of the pillow. Secure the pillow to
the V V e e l l c c r r o o ® on the bottom of the support.
N N o o t t e e
• The head extension with pillow (6100-044-000) is not compatible with the optional equipment hook (6500-147-000),
optional Fowler oxygen bottle holder (6500-241-000), or retractable head section oxygen bottle holder (6085-046-000).
• The head extension pillow is a Type BF applied part.
P P o o s s i i t t i i o o n n i i n n g g t t h h e e o o p p t t i i o o n n a a l l t t h h r r e e e e - - s s t t a a g g e e I I V V p p o o l l e e
C C A A U U T T I I O O N N - - Do not load the IV pole above the safe working load of 25 lb (11.3 kg).
To position the IV pole (Figure 23):
1. Lift and pivot the IV pole from the storage position and push down until the IV pole locks into the receptacle (A).
2. To raise the height of the pole, turn the lock actuator (B) counterclockwise and pull up on the telescoping portion (C) of
the pole to raise it to the desired height.
3. Turn the lock actuator (B) clockwise to lock the telescoping portion in place.
4. For a higher IV pole, pull up on section (D) until the spring clip (E) connects.
5. Hang the IV bags on the IV hook (F).
6. To lower the IV pole, push in on the spring clip (E) and slide section (D) down into section (C). Turn the lock actuator (B)
counterclockwise and slide section (C) into the bottom tube.
7. Turn the lock actuator (B) clockwise to tighten.
8. Lift up and pivot the pole down into the storage position.
N N o o t t e e - - The dual three-stage IV poles (6500-317-000 or 6550-317-000) are not compatible with either the patient right
(6500-315-000 or 6550-315-000) or the patient left (6500-316-000 or 6550-316-000) three-stage IV pole options.
EN
26
6100-109-005 Rev D.1