Filter Instructions - Bovie Smoke Shark III Manual Del Operador

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 83
3.
Install the filter (see filter installation instructions, section 2.3).
4.
The Automatic Activation Device may also be installed by plugging it into the Automatic Activation
Device port located on the front of the Bovie
See section 2.4 for instructions.
5.
Ensure that any smoke evacuation capture accessory is fully installed in the filter port.
6.
Route any other attached accessories to prevent a tripping hazard or crimping of cords, which
could cause unreliable operation.
7.
Activate the Bovie
®
• Pressing the Suction On/Standby Button on the Control Panel of the smoke evacuator.
• Utilizing the automatic activation accessory provided with the unit. This device will synchronize
the activation of the smoke evacuation unit with the activation of the desiccator device.
8.
Deactivate the Bovie
• Pressing the Suction On/Standby Button on the Control Panel of the smoke evacuator.
• Releasing the pencil button on the desiccator device if an automatic activation device
is being used.
9.
Replace the filter when the Filter Life Indicator flashes the color red (0% life remaining). Failure to
change the filter will affect the performance of the Bovie

2.3 Filter Instructions

Bovie
P/N:
®
Configuration:
Bovie
SF06 Filter:
®
Filter(s):
Particle Size, µm:
Filter Life:
Filter Life Indicator:
Filter Installation Instructions:
Note: Before installing or removing any filter, be sure that the Bovie
Evacuator is placed in the Standby mode by pressing the Suction On/Standby Button.
1. Remove the filter from the shipping box and discard any protective wrapping.
2. Inspect the filter for damage, which may have occurred during shipping and storage. Do not install any
filter with visible signs of structural damage.
3. Insert the filter into the filter chamber and ensure the filter is installed completely against the bottom of
the filter chamber.
Filter Removal Instructions:
1. After the filter life has expired, turn the Bovie
mode by pressing the Suction On/Standby Button.
2. Remove all accessories attached to the filter.
3. Remove the Filter from the Bovie
local codes or regulations and facility policy (Section 3.2).
4. Clean the Bovie
Smoke Shark Surgical Smoke Evacuator with a mild disinfectant prior to re-use and
®
follow the indicated instructions for maintenance and installation of a new filter (Section 3.2).
Customer/Technical Services: 800.251.3000 (United States) or +1 615.964.5532 (International)
Smoke Shark Surgical Smoke Evacuator by one of the following:
Smoke Shark Surgical Smoke Evacuator by one of the following:
®
SF06
Wall Mounted
4-Stage Filtration In One Casing, (Pre-Filter, ULPA, Carbon, Post-Filter)
ULPA
0.1 to 0.2 Microns At 99.999% Efficiency
6 hours, Automatic Factory Set Filter Sensor
Time Replacement
Smoke Shark Surgical Smoke Evacuator. Dispose according to your
®
OPERATING INSTRUCTIONS: SECTION 2.0
Smoke Shark Surgical Smoke Evacuator.
®
Smoke Shark Surgical Smoke Evacuator.
®
Smoke Shark Surgical Smoke
®
Smoke Shark Surgical Smoke Evacuator to the Standby
®
P000027200 Rev A Bovie
Page 11
Smoke Shark III Operator's Manual
®