2.6 Electromagnetic Compatibility Information per IEC 60601-1-2

Guidance and Manufacturer's Declaration - Electromagnetic Emissions

The Bovie
Smoke Shark Surgical Smoke Evacuator is intended for use in the electromagnetic
®
environment specified below. The customer or user of the Bovie
Evacuator should ensure that it is used in such an environment.
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Customer/Technical Services: 800.251.3000 (United States) or +1 615.964.5532 (International)
Compliance
Electromagnetic Environment - Guidance
Group 1
The Bovie
energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby
electronic equipment.
Class A
The Bovie
use in all establishments, other than domestic establishments and
those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Class A
Not applicable.
Class A
Not applicable.
OPERATING INSTRUCTIONS: SECTION 2.0
Table 1
Smoke Shark Surgical Smoke
®
Smoke Shark Surgical Smoke Evacuator uses RF
®
Smoke Shark Surgical Smoke Evacuator is suitable for
®
P000027200 Rev A Bovie