Setup and Interconnection

Stryker Endoscopy considers instructional training, or inservice, an integral
part of the 1488 HD Medical Video Camera. Your local Stryker Endoscopy sales
representative will perform at least one inservice at your convenience to help
set up your equipment and instruct you and your staff on its operation and
maintenance. To schedule an inservice, contact your local Stryker Endoscopy
representative after your equipment has arrived.
Setting up the 1488 HD Camera involves three steps:
1. Setting up the console
2. Setting up the camera head
3. Setting up the coupler
Always connect the console to an appropriate power source,
using a hospital-grade power cord. Loss of AC power will cause
the camera to shut down and the surgical image to be lost.
Only connect items to the camera that have been specified for
use with the camera. Connecting incompatible equipment may
cause unexpected results.
When the 1488 HD Camera is used with other equipment,
leakage currents may be additive. Ensure that all systems
are installed according to the requirements of IEC 60601-1
and 60601-1-1.
Equipment which employs RF communications may affect the
normal function of the 1488 HD Camera. When choosing a
location for the 1488 HD Camera, consult the "Electromagnetic
Compatibility" section to ensure proper function.
Always set up the console in a location that allows adequate
ventilation (airflow) to the console. Insufficient ventilation may
cause the console to overheat and shut down.
All manuals and user guides at all-guides.com
EN-11