Classifications
Operating:
(The system may not meet all performance specifications if operated outside
of the following conditions.)
Temperature:
+10 °C to + 40 °C (+50 °F to +104 °F)
Humidity:
20 % to 90 % R.H., non-condensing
Air Pressure:
69 kPa to 110 kPa
Protection Against Electrical Shock: Per IEC 60601-1 (2nd and 3rd Edition
Amendment 1), the Integrated Continuous Battery Charging System is
designed as Class 1 equipment to be used with equipment having Type BF
applied parts.
Flammable Anesthetics: The Integrated Continuous Battery Charging
System is not suitable for use in the presence of a flammable anesthetic
mixture with air, oxygen, or nitrous oxide.
Protection Against Ingress of Fluids: Per IEC 60601-1 (2nd and 3rd
Edition Amendment 1), the Battery Charging System is classified as drip proof
equipment. It is provided with an enclosure that prevents the entry of such an
amount of falling liquid as might interfere with the safe operation of the
injector, indicated by the IPX2 designation.
Mode of Operation: Per IEC 60601-1 (2nd and 3rd Edition Amendment 1)
the mode of operation for the Control Room Unit is continuous with
intermittent loading. The Control Room Unit will operate under normal load for
an unlimited period, without excessive temperature being developed.
The mode of operation for the Scan Room Unit is continuous operation with
intermittent loading. Although power is applied to the Scan Room Unit
continuously, intermittent use for loading and injecting will result in an internal
temperature less than continuous load operating temperatures, but greater
than no-load operating temperatures. Under normal operating conditions with a
minimum of 10 minutes between injections, the internal temperature of the
Scan Room Unit will not rise enough to degrade system performance or
reliability.
3 Specifications
17