3.2. Remote control commands
118
.MeasPt.Z1"6.34" Measured value (SET, MEAS) of the MPList, format de-
pends on the measured quantity
.MeasPt.Z2"25.8" Temperature of the MPList in °C
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP
3.2.2.106. Info.ActualInfo.Display.L1
Info.ActualInfo.Display.L8
Info.ActualInfo.Display.DelAll
Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
1. Lock the display, see 3.2.2.119.
2. Delete the whole display (.DelAll).
3. For writing onto the display, the standard character set will be used.
4. Unlock the display, see 3.2.2.119
5. Delete the whole display (.DelAll).
6. Send a value to nod &Config.Aux.ResDisplay (see 3.2.2.78) to refresh the
display.
$Q sends the contents of the corresponding display line.
3.2.2.107. Info.ActualInfo.Comport.Number
$Q sends the comport number of the Titrino where the PC is connected.
3.2.2.108. Info.Assembly.CycleTime
Info.Assembly.ExV
Inquiries regarding basic variables of the assembly: Cycle time in s, volume of
the active Exchange Unit in mL.
3.2.2.109. Assembly.Bur.Rates.Forward.Selected
Assembly.Bur.Rates.Forward.Digital
Assembly.Bur.Rates.Reverse.Selected
Assembly.Bur.Rates.Reverse.Digital
Expel and aspirating rate.
Digital or analog control. With digital control, the inputted value applies (in
mL/min). "max." means maximum possible rate with the Exchange Unit in
current use.
Analog means rate control with the analog potentiometer on Titrino.
3.2.2.110. Assembly.Bur.Fill
$G starts the 'FILL' mode of the buret function.
3.2.2.111. Assembly.Bur.ModeDis
Assembly.Bur.ModeDis.Selected
Assembly.Bur.ModeDis.V
Assembly.Bur.ModeDis.Time
Assembly.Bur.ModeDis.VStop
Assembly.Bur.ModeDis.AutoFill
up to 32 ASCII characters
up to 32 ASCII characters
$G
read only
read only
read only
digital, analog
0...150, max.
digital, analog
0...150, max.
$G, $H, $C
$G, $S, $H, $C
volume, time
0.0001...0.1...9999
0.25...1...86400
0.0001...9999, OFF
ON, OFF
798 MPT Titrino