ALARM PARAMETERS

The incubator is equipped with several alarm functions.
1. Go to the Tools #1 screen (Menu
Tools#1).
Press "Alarm Setting #1" to display the Alarm
Setting #1 screen. Set automatic temperature
alarm, automatic CO
density alarm, and each
2
alarm delay from this screen (procedure 2).
Press "Alarm Setting #2" to display the Alarm
Setting #2 screen. It is possible to set door alarm
delay, ring back, and remote alarm from this screen
(procedure 3).
2. Enter each parameter on the Alarm Setting #1
screen. Press "Apply" to save the entered value.
The display returns to the Tools #1 screen.
Parameter settings
• Temp. Alarm:
Activated when the chamber temperature varies
from its set value entered here.
o
o
Settable range: ±1.0
C–±5.0
C; factory setting:
o
±1.0
C.
• CO
Alarm:
2
Activated when the chamber CO
density varies from its set value entered here.
2
Settable range: ±0.5 %–±5.0 %; factory setting: ±1.0 %.
• Alarm Delay:
When the conditions for the automatic temperature alarm or the automatic CO
density alarm are reached,
2
the alarm buzzer will sound after the time set here has elapsed.
Settable range: 0 minutes
15 minutes; factory setting: 15 minutes.
Note: When the conditions return to the normal range within the alarm delay time, the buzzer does not
sound.