Mestra R-100138 Manual Del Usuario página 13

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

OPERATION
Remove the cover and put it on the cover holder vertically. Put items in the
basket, then put them in the cleaning tank. Add water to a level between
MIN and MAX and above the area to be cleaned.
If the unit is turned on without water, ultrasonic energy will not be absorbed.
This model also has high power. Once on for over 15 seconds, it may
damage the unit or severely reduce the life of the unit.
Connect the power cord (L) and turn the switch on (I). LED display shows
. This is the most common timer setting and working state. If the timer
needs to be adjusted, press
30 minutes. Common cleaning times are 5, 10 or 15 minutes.
Prolonged cleaning time may result in:
a. Loosening of the screws if used.
b. Increasing pre-existing cracks.
c. Peeling of coating which was already separated.
Press
to start cleaning.
Durring cleaning, buzzing sound can be heard from the cleaning tank,
indicating that the cleaning is underway. Closing the lid will reduce the noise
level.
The digital timer will count down to show the remaining cleaning time. When
it displays
, the cleaning is done.
To stop cleaning at any time, press
The unit is designed with overheating protection to avoid extended and
continuous operation for too long. When the red light is on, the unit will not
turn on even if it is forced to. The unit will be idled for about 20 minutes and
can be restarted after the green light comes on.
When the heater is hended, press
temperatures. Press
turned on. When the water reaches the set temperature,
the heater is turned off.
To ensure safety, the unit is designed with dual protections. The heater will
be turned off automatically if it has been running for 50 minutes.
If solution is added, to speed up the mixing and to use less solution, use
degas function. Press
Function button
, The icon
changes over to normal cleaning automatically after 90 seconds.
When debris "smoke" is no longer visible, cleaning is done. If additional
cleaning is needed, reset the timer and repeat the steps above.
When cleaning is completed, switch off the power, open the lid and retrieve
the basket and the items.
Alternatively, raise and suspend the basket above water to drain water.
Connect the hose to the drainage pipe as shown.
Open the drainage valve to drain the dirty water.
Clean and wipe the tank dry.
Close the drainage valve.
button. The timer can be set between 1 to
button.
button to select one of the 5 preset
button, the icon
flashes, indicating the heater is
button tos et the timer to 10 minute. Then press
illuminates. Press
dims, indicating
to start degassing. It
13