Characteristics Of The Water To Be Used; Loading Of The Materials Into The Autoclave - Tecno-gaz ONYX B 5.0 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 173
10.3 Characterisstics of the water to be used
TABLE SHOWING THE QUALITY LEVELS LAID BY THE EN13060
Evaporation residue
Silicon oxide
Iron
Cadmium
Lead
Remains of heavy metals apart from iron, cadmium,
lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20°C)
Ph value (acidity level)
Appearance
Hardness (and alkaline earth ions)
10.4 Loading of the materials in the autoclave
Arranging the materials to be sterilized on the provided trays, as follows:
• Do not superpose the materials
• Arrange the wrapped materials with the paper side facing upwards
• Never bring the materials into contact with the sterilization chamber or the autoclave door
• Put scissors and dental forceps with blades open
After loading the materials, close the door of the autoclave. The display will show the icon
and the message DOOR CLOSED.
CEN STANDARD EN13060

10

1

0.2

0.005

0.05

0.1

2

0.5

15
5  7,5
Transparent, clear, without deposits

0.02
ENGLISH
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
s/cm
mmol/l
19

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Onyx b 8.0Ab015zxyAb017zxz

Tabla de contenido