MINIASPEED BATTERY PRO is intended for use in the elctromagnetic environment specified below. The customer
or the user of the device should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
EN 61000-4-2
Burst/Fast Transient
EN 61000-4-4
Surge
EN 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
EN 61000-4-11
Power frequency
magnetic field
EN 61000-4-8
MINIASPEED BATTERY PRO is intended for use in the electromagnetic environment specified below. The customer
or the user of the navigator should assure that it is used in such an electromagnetic environment.
Immunity test
RF conducted
EN 61000-4-6
RF radiated
EN 61000-4-3
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survery, c should be less than
the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the
following symbol:
Test level
EN 60601-1-2
± 6kV contact
± 8kV air
±2kV power supply lines
±1kV differentail mode
±2kV common mode
< 5% UT
(>95% dip in UT)
for 0,5 cycles
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(>95% dip in UT)
for 5 seconds
3 A/m
Immunity ascpetcts at r.f.
Test Level
EN 60601-1-2
3 Veff from 150kHz
to 80MHz
3 V/m from 80MHz
to 2,5GHz
Immunity aspects
Compliance Level
± 6kV contact
± 8kV air
±2kV power supply lines
±1kV differential mode
±2kV common mode
< 5% UT
(>95% dip in UT)
for 0,5 cycles
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(>95% dip in UT)
for 5 seconds
3 A/m
Compliance Level
3 Veff from 150kHz
to 80MHz
3 V/m from 80MHz
to 2,5GHz
Electromagnetic environment -
guide
Floors schould be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the device
requires continued operation during
power mains interruptions. It is
recommended that the Laser/Pulsed
Light BiosYag be powered from an
uninterruptible power supply or a
battery.
Magnetic power frequency fields
should be that of a typical commercial
or hospital environment.
Electromagnetic environment -
guide
Portable and mobile RF
communications equipment should
be used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from that equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1,2 ◊÷P da 150kHz a 80MHz
d = 1,2 ◊÷P da 80 MHz a 800 MHz
d = 2,3 ◊÷P da 800 MHz a 2,5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).