H y b r i g e n e O P E R A T O R ' S M A N U A L
Warning
THE UNIT WORKS UP TO 80°C. This temperature can cause burns. Techne
take great efforts to protect operators from hazards, but the following points are
brought to your attention:
1
Use care and wear protective gloves to remove hybridisation tubes from the cabinet
if the unit is at high temperature.
If the protocol allows, it is better to switch off the unit and open the door to allow the
unit to cool to below 40
2
The coloured end cap of the 80mm large tube and the 44mm small tube has two
holes so you have to be careful not to spill the probe through these.
One is the hole which allows the air to expand when the tube is heated.
The other is the filling hole which is revealed when the spindle cap is removed.
DO NOT TIP THE TUBE SO THAT THE COLOURED END IS DOWN AS THIS
WILL CAUSE A SPILLAGE.
3
The unit has been designed to conform to international safety specifications and is
fitted with a resettable over-temperature cutout. If, for any reason, the temperature
rises above a safe level for the instrument, the overtemperature protection cuts out
the heater, but the tubes continue to rotate. Check the reason for the cut out and, if
there is no obvious cause, reset the cutout once the unit has cooled. The reset, on
the back of the unit, is operated by pressing the centre of the button.
4
If the Controller displays 'Err', the thermistor has broken or become disconnected
and the unit will cool down automatically.
Servicing by Techne and their Agents
Please note that neither Techne nor any of their agents can accept the return of
ANY GLASSWARE from these units.
If the unit is returned it must be thoroughly decontaminated and a CERTIFICATE
OF CLEANLINESS returned with the unit.
A copy of the certificate is available from our web-site, www.techne.com, or can
be requested from the Service Department if the unit is being returned.
16
C before attempting to remove the tubes.
o