9.2 ERROR MESSAGES ON SLEEPSTYLE SCREEN

If a fault is detected with your device or its accessories, an error message will appear on the display screen. Identify the error
code in the ranges specified below and follow the appropriate corrective action. If the error persists or reoccurs, please consult
your healthcare provider. Do not attempt to repair the device yourself.
Error codes between
100–199
400–499
510 or 512
500–599 (excluding 510 or 512)
SleepStyle, SensAwake, ThermoSmart, and F&P InfoUSB are trademarks of Fisher & Paykel Healthcare Ltd.
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG Inc. and any use of such marks by
Fisher & Paykel Healthcare is under license. Other trademarks and trade names are those of their respective owners.
Android and Google Play are trademarks of Google Inc.
"Made for iPhone" means that an electronic accessory has been designed to connect specifically to iPhone, and has been
certified by the developer to meet Apple performance standards. Apple is not responsible for the operation of this device or
its compliance with safety and regulatory standards. Please note that the use of this accessory with iPhone may affect wireless
performance. Apple, OS X, Mac and iPhone are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is
a service mark of Apple Inc., registered in the U.S. and other countries.
Windows is either a registered trademark or trademark of Microsoft Corporation in the United States and/or other countries.
For patent information, see www.fphcare.com/ip.
Solution
Your device may not be able to provide effective therapy. Your device may have shut
down or may not be able to provide your prescribed pressure.
Solution: Turn the power off and on at the power supply to restart the device.
Humidity may have been disabled. Your device is still safe to use without humidity. You
will still be treated at your prescribed pressure.
Solution: Turn the power off and on at the power supply to restart the device.
There may be a problem with your ThermoSmart breathing tube. Your device is still safe
to use. You will still be treated and get humidity, but it may not be optimal.
Solution: Try reconnecting your ThermoSmart breathing tube. When connected correctly,
the ThermoSmart icon
will appear on the home screen. Alternatively, turn the power
off and on at the power supply to restart the device.
The ThermoSmart breathing tube may have been disabled. Your device is still safe to use.
You will still be treated and get humidity, but it may not be optimal.
Solution: Turn the power off and on at the power supply to restart the device.