8-4 Lubrication
NSK PANA SPRAY Plus
Apply NSK PANA SPRAY Plus every time after each use and/or before autoclaving.
1) Remove the bur from the handpiece.
2) Insert the Spray Nozzle into the Spray Port nozzle on the can.
3) Insert the Spray Nozzle in rear of the handpiece. Hold the handpiece and spray for approximately 2 - 3
seconds. Apply lubricant until it expels from the handpiece head for at least 2 seconds (Fig. 8).
CAUTION
• Debris and blood could enter the product during surgical operations. To avoid accumulation of debris inside the product, NSK recommends to
lubricate the product with PANA SPRAY Plus immediately after use and before sterilization.
• When applying spray be sure to hold the handpiece firmly to prevent the handpiece from slipping out of the hand due to the spray pressure.
• Hold the spray can upright.
12
Spray Nozzle
Handpiece
Fig. 8