Preparation for use
To be used by healthcare professionals or adequately trained caregivers only.
Checklist
Prior to using the Invia Motion NPWT system, it is important to check for the following:
– Completeness and general condition of the packaging
– Damage to the power cord and plug
– Damage to the pump
– Proper functioning of the device
– Battery status
Connect Invia Motion canister/tubing set 0.15 l (150 ml)
Make sure the correct canister is used:
4. Connect the dressing tubing to the pump tubing.
Make sure both parts are aligned correctly as shown in the picture below. Push together until you hear
a click.
For dressing applications, please refer to the instructions for use
provided with the Invia dressing kits.
14
1. Open sterile packaging.
CAUTION
Do not use the Invia Motion canister/tubing set if the sterile packaging is
damaged or has expired.
2. Insert the Invia Motion canister/tubing set into the Invia Motion NPWT pump
connecting port (1). Rotate the canister/tubing set to the right to secure (2).
3. Make sure the canister is aligned with the label to the front.
087.4016.