WARNING:
•
Always switch off the device, disconnect it from any voltage source or power
supply, and remove the test probes. Otherwise there might be the danger of
damaging the device or personal injury.
•
Only continue using the device, after the putting everything back together
according to the instructions.
Replacing Fuse
1.
Turn off the power supply of the instrument, and remove the probes from the input
sockets.
2.
Remove the protective rubber casing. Use screwdriver to unscrew the screws fixing
the back cover and remove the back cover.
3.
Remove the burnt fuse, replace it with a new fuse of the same specifications (mA:
F600mA/250V fuse, size: 6*32mm; 10A: F10A/250V fuse, size: 6*32mm), and
ensure that the fuse is clamped in the safety clip.
4.
Install the back cover, fix and lock it with the screws. Put back the protective rubber
casing.
46