Storz 11272 V Serie Manual De Instrucciones página 65

Tabla de contenido

Publicidad

63
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
The flexible CMOS Video-Cysto-Urethroscope 11272 V/VU in conjunction with monitor 8402 ZX is intended for use in the electromagnetic environment
specified below. The user should ensure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
acc. to IEC 61000-4-2
± 8 kV air
± 2 kV for power supply
Electrical fast transient/burst
lines
acc. to IEC 61000-4-4
± 1 kV for input/output lines
± 1 kV differential mode
Surge acc. to IEC 61000-4-5
± 2 kV common mode
<5% U
(>95% dip in U
for 0.5 cycle
40% U
T
(60% dip in U
Voltage dips, short
interruptions and voltage
for 5 cycles
variations on power supply
input lines acc. to IEC 61000-
70% U
T
4-11
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
Power frequency (50/60
Hz) magnetic field acc. to
3 A/m
IEC 61000-4-8
is the a.c. mains voltage prior to application of the test level.
* Note: U
T
Appendix
Electromagnetic Compatibility (EMC)
Information
Table 202
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance level
Complies
± 6 kV contact
± 8 kV air
Complies
± 2 kV for power supply lines
± 1 kV for input/output lines
Complies
± 1 kV differential mode
± 2 kV common mode
Complies
<5% U
*
T
*
(>95% dip in U
)
T
T
)
for 0.5 cycle
T
Complies
40% U
T
)
(60% dip in U
)
T
T
for 5 cycles
Complies
)
70% U
T
T
(30% dip in U
)
T
for 25 cycles
T
)
Complies
T
<5% U
T
(>95% dip in U
)
T
for 5 sec
Complies
3 A/m
Anexo
Indicaciones sobre compatibilidad
electromagnética (CEM)
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Mains power quality should be that of a typical commercial
or hospital environment. If image distortion occurs it may be
necessary to further filter the power quality by using an AC
line filter.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
If the user of the equipment or system requires continued
operation during power mains interruptions, it is
recommended that the equipment or system be powered
from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

11272 vu serie11272 vk serie11272 vuk serie8402 zx

Tabla de contenido