Integra CUSA Excel Instrucciones De Uso página 4

Ocultar thumbs Ver también para CUSA Excel:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 11
EN
EN
4. Remove the o-rings.
5. Locate the slot in the tip torquing base that
matches the color on the handpiece connector
(23 kHz – green; 36 kHz – blue). Put the
handpiece in the tip torquing base so that the
metal connecting body fits snugly in the metal
end of the slot.
6. Hold the handpiece in place in the tip torquing
base.
Notice
To avoid product damage, NEVER hold the
handpiece in your hand while using the torque
wrench to tighten or loosen the tip.
7. Match the colored end of the torque wrench with
the handpiece connector color. Slide the color
coded side of the wrench over the tip, being
careful not to damage the tip, until the hex in the
wrench engages the hex of the tip.
8. Rotate the wrench counterclockwise until the tip
is loose.
9. Carefully remove the wrench from the tip.
10. Remove the handpiece from the tip torquing
base.
11. Unscrew the tip.
4
12. Discard the flue and o-rings following hospital
protocols for biohazardous waste.
Tip Torquing Set Cleaning
Notice
Product damage will result if you do not follow
these notices when cleaning the handpiece:
Do not use ultrasonic or automated washers.
Do not autoclave.
Do not use chlorinated substances such as
bleach solution.
Do not clean with abrasives such as Ajax,
Comet or steel wool.
1.
Using a soft cloth and mild cleansing solution
or blood dissolving detergent, manually clean
the tip torquing base and the torque wrench.
2.
Rinse the tip torquing base and the torque
wrench thoroughly with water.
3.
Using a soft cloth, dry the tip torquing base
and torque wrench.
For instructions on cleaning the handpiece, refer to
the CUSA Excel System User's Guide.

Publicidad

Tabla de contenido
loading

Tabla de contenido