IN SITU MONITOR USABILITY TEST

We recommend performing the usability test daily, but at least when
changing the location of the bed or monitor.
• Make sure the green light flashes while the baby is in the cot. The green light reacts
by flashing with the baby's breathing or movements. The flashing of the light may
not be regular - the frequency corresponds to the baby's movements or breaths.
• Then take the baby out of the cot and move away from the cot. Wait a while for the
vibrations from your movements and the mattress to subside.
• If a pre-alarm sounds after 17 secs and an alarm after 20 secs, the monitor has
passed the usability test and can be fully relied upon. Verify that the alarm sound
can be heard in all areas where the parents or other caregivers are present.
If the green light flashes even when the baby is not in the crib, the
monitor is detecting disturbance in the surroundings. Surrounding
shocks or vibrations with a similar frequency can be falsely evaluated
by the device as the baby's breathing / movements, therefore they
must be eliminated in order to ensure the device works reliably and
the baby is safe! Disturbance can be caused by intense airflow
(fans, air conditioning), walking near the bed, mechanical vibration
from household appliances, etc. Eliminate disturbance in the
surroundings or relocate the cot!
5.

POWER SUPPLY AND REPLACING THE BATTERIES

The device is powered by two alkaline 1.5 V/AA batteries and monitors their power.
The device distinguishes between low and critical battery level.
Low battery is indicated by a flashing red light with a picture of a battery. All
functions except the lamp are retained. The low battery indication lasts for about 2
weeks before the batteries are completely empty, so you have enough time to replace
them. The batteries must be replaced as soon as possible after the red low battery
indicator comes on.
In the event of a critically low battery, the fault is indicated audibly during the
automatic test and the device does not turn on. Replace both batteries immediately!
The batteries need changing depending on the intensity of use - usually after 4-12
months. Before replacing the batteries, switch off the device by pressing
the mechanical button. Remove the battery cover and remove the original batteries.
Remove the batteries when you are not using the monitor at all.