EN | Baby breathing monitor
6.1
LIFE SPAN OF THE SENSOR PAD
The life span of the sensor pad is set at 2 years, after which the pad must
be replaced.
As auxiliary data, the Nanny BM-03 breathing monitor also monitors the number of
hours the monitor has been in operation - if it is likely the pad has exceeded the
specifi ed life span, a warning beep will sound 3 times during each automatic test.
However, the date it is put into service is decisive for determining the life span.
The sensor in the pad may wear out over time, so it may not correctly sense your baby's
movements and breathing. In this case, the monitor may give false alarms that will
cause you unnecessary worries. False alarms may not endanger your baby's life
or health, but you will hear the alarm so often that you prefer to turn off the monitor
or reduce your alertness, which could endanger your child.
7.
MAINTENANCE AND CLEANING
The device does not require any special maintenance other than replacing the batteries
and cleaning. We simply recommend that you check the sensor pad in the cot from
time to time to ensure moisture has not condensed where the pad touches
the mattress. It is advisable to turn the mattress 180 ° in the cot once in a while,
or turn it upside down or let it air, etc. Only use a slightly damp cloth for cleaning.
Avoid wet wipes or other materials that may have loose fi bres that can clog the holes
in the control unit.
It is not necessary to sterilize the breathing monitor when in medical facilities.
The device does not require sterilization or disinfection between use by different
patients. The device can be treated with commonly available disinfectants at
a concentration in line with the manufacturer's recommendations for the given level
of disinfection. Keep in mind that moisture entering the battery compartment
or the control unit during cleaning can damage the device. Only use a cloth
dampened with disinfectant solution for disinfection and only wipe the device lightly.
The frequency of cleaning is not specifi ed by the manufacturer and the frequency
of cleaning does not affect the product's life span. As concerns further
maintenance, the device does not have a measuring function to be calibrated.
During maintenance, protect the pad, supply cable and connectors
from mechanical damage and moisture entering.