EN | Baby breathing monitor
12.
TECHNICAL SPECIFICATIONS
Power supply
Idle consumption
Consumption during an alarm
Low battery indicator voltage
Critically low battery indicator voltage
Threshold breathing rate
Measuring range for room temperature
Typical battery life - home health care
Typical battery life - health care provider
Sensor pad
Acoustic alarm level
Electronic control unit - dimensions
Operating conditions
Transport and storage
Ingress protection
Life span of the sensor pad
Service life of the control unit
Service life of the device
Meets the following standards:
13.
DEFINITIONS AND SYMBOLS
ME device
Medical electrical device (with attachment that detects energy from the patient)
EMC
Electromagnetic compatibility (a set of normative requirements for a ME device)
SIDS
Sudden infant death syndrome
Designation of the attachment part type BF
General sign for compulsory activity
General warning
Refer to the operating manual
IP31
The device's immunity to the ingress of foreign bodies and water (Ingress protection class)
Product identifi cation by barcode
Prefi x: 0xxxy product identifi cation (02594 unit; 02593 pad): Suffi x: nnnnnnnn serial number
0xxxxynnnnnnnn
20
MXA51105
3 V; 2 × 1.5 V alkaline battery type AA (LR6)
106 μA
270 mA
2.46 V ± 0.15 V
2.2 V ± 0.15 V
< 8 breaths/min (i.e. < 0.13 Hz)
-40° to 85°C with an accuracy of ± 0.2°C
6 months (this is reduced with frequent alarm testing and using
the lamp)
4 months (this is reduced with frequent alarm testing and using
the lamp)
Type BM-03D, dimensions max. 300 × 500 × 15 mm
80 dB/m ±5 % dB/m
max. 140×82×37 mm, weight 125 g + batteries
+5 °C to +40 °C, RH 15 % - 93 %, 700–1200 kPa
0 °C to +70 °C, RH 10 % - 85 %, 700–1200 kPa
IP31
2 years from putting into service
10 years
10 years under the condition of replacing the sensor pad after its
life span (i.e. every 2 years)
EN 60601-1:2006/A1:2012/Cor.1:2014
EN 60601-1-2:2015
EN 60601-1-6:2010/A1:2015
EN 60601-1-11:2015
EN 62366-1:2015; EN 50581:2012
EN ISO 10993-1:2009/AC:2010
EN ISO 14155:2011; EN 14971:2019
EN 14485:2003