The alarm can be turned off by pressing the mechanical on/off button.
The monitor operates in active and standby mode. In active mode, the monitor detects
the baby's breathing movements and additional functions are activated. In standby
mode, the monitor is switched to power saving mode and the unit evaluates the signals
from the pad to be able to draw attention to a baby being placed there.
A quiet beep indicates switching to standby mode.
If the button is constantly pressed it is a defect. If this fault is detected during the
automatic test after switching from standby mode, the control unit will not start. If a
mechanical button fault is detected in active mode, a critical failure will be reported
visually and acoustically. Release the mechanical button to remove the fault.
Automatic functionality test
When switching from standby to active mode or when inserting the batteries, the
device automatically tests its functionality. The test checks the batteries, the correct
type of pad is connected and its age, the optical and acoustic signal, the notifi cation
for the on/off switch being in the incorrect state for the baby's current absence or
presence.
Functionality test results:
A) All indicators fl ash and short beep = all checks were successful and the device
is fully functional.
B) Flashing lights and repeated warning beeps = error diagnosed but it does not
prevent the monitor from being used and maintaining its functionality.
None – the device is fully functional
The Switch Guard function is deactivated
Connected a sensor pad or control unit with possibly
exceeded life span
C) Repeated warning beeps and the device does not switch on = a critical fault has
been detected (critically low battery), the device cannot be used. Correct the error
(insert new batteries), then the device can be used again. If the error persists, send
the monitor for a service.
2.2

SUPPLEMENTARY FUNCTIONS

2.2.1
DETECTING WHETHER THE BABY HAS BEEN PLACED ON THE SENSOR PAD OR
TAKEN OFF IT (SWITCH GUARD)
When in standby mode the BM-03 Baby Breathing Monitor continuously evaluates
signals from the pad and is thus the only device on the market today to point out that
the baby was most likely placed on the sensor pad. This prevents tragic consequences
if a parent or other caregiver forgets to turn on the device and the baby stops
breathing.
It can also warn parents or other caregivers if they remove a baby from the cot and
forget to turn off the device.
.
Type of warning
Acoustic indication
1x beep
2x beep
3x beep
MXA51105
7