5.7 Measurement of resistance
WARNING
Do not perform measurements on the circuit under the
voltage. Before the measurement disconnect the pow-
er and discharge capacitors.
To perform measurement of resistance:
set the rotary switch at Ω
connect black test lead to COM terminal, and red test lead to
VΩLoZV
contact the tips of test probes to the points of measurement; the
best solution is to disconnect one side of the tested element, to
prevent the remaining part of the circuit interfere with the read-out
of the resistance value,
read the measurement result on the display.
5.8 Circuit continuity test
WARNING
Do not perform measurements on the circuit under the
voltage. Before the measurement disconnect the pow-
er and discharge capacitors.
To perform the continuity test:
set the rotary switch at Ω
connect black test lead to COM terminal, and red test lead to
VΩLoZV
press MODE button, to display Ω
contact the tips of test probes to the points of measurement,
read the measurement result on the display; the beep will be acti-
vated when resistance values are below approx. 50 Ω.