Replacement Procedure - Chauvin Arnoux CA 6161 Guia De Inicio Rapido

Ocultar thumbs Ver también para CA 6161:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 35
FUSES
FUSE F1
Fuse F1 protects the instrument in continuity measurement with a high
current (10 or 25 A).
To check F1:
Put the instrument in continuity measurement,
Select the External Terminals
Make a short circuit by connecting a safety lead between terminals C1 and C2.
For configuration, choose a measuring current of 10 A and a measurement of 2 wires
Press the Start / Stop button to start the measurement.
If current I is close to 0, fuse F1 is faulty.
FUSES F2 AND F3
Fuses F2 and F3 protect the power supply to the instrument.
To check F2 and F3:
Connect the mains lead between the instrument socket and the mains.
Press the On / Off switch. The instrument starts.
If the instrument does not start, one of the two fuses F2 or F3, or both fuses are defective. Either way, replace both fuses.

REPLACEMENT PROCEDURE

Disconnect anything connected to the instrument and switch it off.
For F1, press the fuse holder while unscrewing it a quarter turn.
For F2 and F3, unscrew the fuse holder a quarter turn using a flat screwdriver.
Remove the defective fuse and replace it with a new fuse.
To ensure continuity of safety, replace a defective fuse only with a fuse with strictly identical characteristics.
F1: FF 30 A 300 V 6.3 x 32 mm
F2 and F3: FF 16 A 300 V 6.3 x 32 mm
Close the fuse holder by tightening it a quarter turn. For F1, press while tightening.
Check the correct operation of the replaced fuse as described above.
then
connection.
18
F1
30 A 300 V
16 A 300 V
6,3 x 32
.
F2
F3
16 A 300 V
6,3 x 32
6,3 x 32
.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ca 6163

Tabla de contenido