8. Troubleshooting
Malfunction
The device cannot be switched on.
The device cannot be switched on.
The suction intensity is decreased.
The suction intensity is decreased.
Liquid or mucus got into the
connecting tube.
Liquid or mucus got into the
connecting tube.
Liquid or mucus got into the
connecting tube.
9. Important information about
electromagnetic interferences
Home wireless network equipment, mobile phones and wireless
phones may affect the operation of certain medical equipment,
as electromagnetic interference can occur between devices.
Electromagnetic interference may cause harmful interference
in the operation of certain medical devices, potentially causing
emergency situations.
User manual * DMR-01/A05/V08 * 2022-08-07 * Nosiboo Pro Nasal Aspirator * Page: 14/52
Reason
The plug is not connected.
The device automatically switched off
for safety reasons due to overuse.
The nose tip is blocked or the Colibri head
was incorrectly assembled.
The air mesh is clogged.
The Colibri head was not held in the
correct position.
The connection point of the tube is not
pointed downwards.
Too much saline solution was used.
Due to the above mentioned reasons, medical devices must not
interfere with any other equipment.
Do not use near high-performance radiation-emitting devices.
The Nosiboo Pro Nasal Aspirator is intended for use in an electro-
magnetic environment in which radiated RF disturbances are
controlled.
The Nosiboo Pro Nasal Aspirator has been tested and found to
comply with the requirements of EN 60601-1-2:2007 and of FCC
Part 15, 2007.
Recommendation
Connect the power cord to the device,
then to the 110 V socket.
For the sake of your child's health, stop
using the nasal aspirator. The device can
be used again in approx. 45 minutes.
Take the Colibri head apart and clean it.
After drying the parts of the Colibri head,
assemble them according to the user manual.
Replace the air mesh according to the
user manual's "Cleaning, maintenance
and storage" section.
Switch off the device and clean and dry
the Colibri head and the connecting tube.
Check the air mesh and replace it if necessary.
Switch off the device and clean and dry the
Colibri head and the connecting tube.
Check the air mesh and replace it if necessary.
Reduce the volume of the saline solution to
less than a teaspoon (0.17 oz, 5 mL). Switch
off the device and clean it, then dry the Co-
libri head and the connecting tube. Check
the air mesh and replace it if necessary.