User Keys - Gima DIATERMO 106 Manual De Instrucciones

Ocultar thumbs Ver también para DIATERMO 106:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

GIMA SpA
DIATERMO 106
4.3.1 U
K
SER
EYS
The keys are split into two groups, the four operative keys, positioned at the left of the unit, and the two output power control keys
under the digital display.
The two output power control keys are used to increase (+) or decrease (-) the output current.
The function of each of the operative keys is as follows:
Pressing this key Cut current, i. e. pure sine current is preset, to be delivered when the foot-switch output control is activated. Crest
factor of the Cut current is 1.5. Cut current preset condition is indicated by green LED which will illuminate.
Pressing this key Blend current, i.e. modulated current is preset, to be delivered when the foot-switch output control is activated.
Crest factor of the Blend current is 1.8. Blend current preset condition is indicated by green LED which will illuminate.
Pressing this key presets Coag current, i.e. modulated current is preset, to be delivered when the foot-switch output control is
activated. Crest factor of the monopolar Coag current is 2.5. Coag current preset condition is indicated by green LED which will
illuminate.
29
Manuale d'Istruzione / Instruction's Manual / Manual de Instrucciones
MA126IGBa